






| Cat. no. | Capacity | Build | Controller |
|---|---|---|---|
| HTN-AV365 | 75 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV366 | 75 L | Prime (All SS 304) | LED (standard) |
| HTN-AV367 | 115 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV368 | 115 L | Prime (All SS 304) | LED (standard) |
| HTN-AV369 | 150 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV370 | 150 L | Prime (All SS 304) | LED (standard) |
| HTN-AV371 | 175 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV365-G | 75 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV365-R | 75 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV366-G | 75 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV366-R | 75 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV367-G | 115 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV367-R | 115 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV368-G | 115 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV368-R | 115 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV369-G | 150 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV369-R | 150 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV370-G | 150 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV370-R | 150 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV371-G | 175 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV371-R | 175 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV372 | 175 L | Prime (All SS 304) | LED (standard) |
| HTN-AV372-G | 175 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV372-R | 175 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV373 | 235 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV373-G | 235 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV373-R | 235 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV374 | 235 L | Prime (All SS 304) | LED (standard) |
| HTN-AV374-G | 235 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV374-R | 235 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV375 | 339 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV375-G | 339 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV375-R | 339 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV376 | 339 L | Prime (All SS 304) | LED (standard) |
| HTN-AV376-G | 339 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV376-R | 339 L | Prime (All SS 304) | Recording (21 CFR software) |
| HTN-AV377 | 430 L | Classic (SS 304 inside, SS 202 outside) | LED (standard) |
| HTN-AV377-G | 430 L | Classic (SS 304 inside, SS 202 outside) | Graphic LCD (F0 value) |
| HTN-AV377-R | 430 L | Classic (SS 304 inside, SS 202 outside) | Recording (21 CFR software) |
| HTN-AV378 | 430 L | Prime (All SS 304) | LED (standard) |
| HTN-AV378-G | 430 L | Prime (All SS 304) | Graphic LCD (F0 value) |
| HTN-AV378-R | 430 L | Prime (All SS 304) | Recording (21 CFR software) |
Downloads
A high pressure, triple walled horizontal autoclave built into a flat, square stainless steel front panel, from 75 to 430 L. The door, gauges and controller sit flush on the panel, so the unit can stand against a wall or be built into a partition, with the cylindrical chamber and its stand behind. Steam fills the jacket around the chamber first, so the chamber walls stay hot: cycles start faster, less steam condenses on the load, and with the moisture trap on the discharge line the load comes out dry. It runs at up to 30 psi, so it sterilizes at 134 °C for fast cycles as well as at 121 °C at 15 psi. The working pressure is adjustable from 15 to 30 psi, and every unit is hydraulically tested to 2.5 times its working pressure.
Choose in three steps: the size, the build and the controller. Classic has an SS 304 chamber and jacket with an SS 202 front panel, on a black powder coated mild steel stand. Prime is all SS 304, stand included. The stainless steel door closes with a radial lock, and the self-locking safety door cannot be opened until the chamber pressure is released. A separate SS boiler (steam generator) comes as standard, and a multiport valve sends steam to the jacket and the chamber and exhausts them. An automatic pressure control switch holds the set pressure, a vacuum breaker breaks any vacuum formed as the steam condenses, and an automatic low water cut-off protects the heater. Glass wool insulation keeps the heat in. It works on a 440 V three phase supply.
Only need 121 °C? See the standard version, HTN-AV311.
For laboratory use only.
Features
- Steam jacket keeps the chamber hot for faster, drier cycles
- LED, Graphic LCD or Recording controller runs the full cycle
- High pressure: 134 °C at 30 psi, or 121 °C at 15 psi
Build materials
| Part | Classic | Prime |
|---|---|---|
| Inner chamber | SS 304 | SS 304 |
| Steam jacket | SS 304 | SS 304 |
| Front panel | SS 202 | SS 304 |
| Door | SS 304, radial lock | SS 304, radial lock |
| Stand | MS, black powder coated | SS 304 |
How the cycle runs
- Heat up: steam fills the jacket first, then the chamber, and air is let out.
- Sterilize: the timer runs at the set temperature.
- Cool down: a buzzer sounds, the heater stops and the pressure falls.
All three controllers run this cycle by themselves. The timer starts only at the set temperature, so every load gets its full time. The cycle stops if the water runs low or the temperature goes too high, and the pressure control switch cuts the heater if the pressure rises too high.
Compare controllers
| Feature | LED | Graphic LCD | Recording |
|---|---|---|---|
| Runs the full cycle automatically | ✓ | ✓ | ✓ |
| Temperature and pressure display | ✓ | ✓ | ✓ |
| Stage and time left on one screen | – | ✓ | ✓ |
| F0 value: how much sterilizing each load received | – | ✓ | ✓ |
| Choose when the cycle ends: as soon as sterilizing time is over, once the temperature has fallen by a set amount, or after a set cooling time | – | ✓ | ✓ |
| Four sensors map the load | – | – | ✓ |
| PC report software built for 21 CFR compliance | – | – | ✓ |
| Batch records and reports on a PC | – | – | ✓ |
| Printer for a paper record (option) | – | – | ✓ |
Why upgrade?
- Graphic LCD: see the whole cycle at a glance. The F0 value adds up the sterilizing effect of every minute above 100 °C, heat up and cool down included, so you can see how much each load received.
- Recording: every batch is recorded and reported on a PC with software built for 21 CFR compliance, which supports audits and validation. Four sensors check the load itself, not just the chamber.
What is the F0 value?
- What: the sterilizing effect of a cycle, counted in minutes at 121 °C.
- How: the controller reads the chamber temperature all through the cycle. A minute at 121 °C counts as 1, at 111 °C as 0.1, at 131 °C as 10. It adds them up, heat up and cool down included.
- Why: the same timer setting can give a big or slow-heating load less sterilizing than a small one. F0 shows what each load actually got.
- Reading it: F0 15 means the same effect as 15 minutes at 121 °C. If a load ends below the F0 your method needs, run it again or check the load.
Please note: F0 shows how much sterilizing a load received, not whether it is sterile. Only biological indicators placed in the load can confirm that.
Sizes
| Capacity | Chamber (dia. × depth) | Heater load | Classic Cat. no. | Prime Cat. no. |
|---|---|---|---|---|
| 75 L | 400 × 600 mm | 6 kW | HTN-AV365 | HTN-AV366 |
| 115 L | 400 × 900 mm | 6 kW | HTN-AV367 | HTN-AV368 |
| 150 L | 400 × 1200 mm | 9 kW | HTN-AV369 | HTN-AV370 |
| 175 L | 500 × 900 mm | 9 kW | HTN-AV371 | HTN-AV372 |
| 235 L | 500 × 1200 mm | 12 kW | HTN-AV373 | HTN-AV374 |
| 339 L | 600 × 1200 mm | 12 kW | HTN-AV375 | HTN-AV376 |
| 430 L | 600 × 1500 mm | 12 kW | HTN-AV377 | HTN-AV378 |
Add -G for the Graphic LCD or -R for the Recording controller, for example HTN-AV365-G.
Options
- Vacuum pump for pre- and post-vacuum drying
- Double door (pass-through) version, for a clean side and a dirty side
- Thermal printer with the Recording controller
Options are quoted on request.
Performance
- Pressure15 to 30 psi working (1.05 to 2.1 kgf/cm²); hydraulically tested to 2.5 times working pressure
- HeatingFlanged immersion heating elements, protected by low water cut-off
- Temperature controlAutomatic cycle controller in three versions: LED, graphic LCD with F0 value, and recording with PC reports
- Temperature range121 °C and 134 °C cycles, 134 °C max.
General
- Inner chamberSS 304
- InsulationGlass wool between outer chamber and steam jacket
- DoorRadial locking lid with jointless silicone rubber gasket
- SafetyAutomatic pressure control, spring loaded safety valves on boiler and chamber, dead weight release valve, automatic low water cut-off, self-locking door, vacuum breaker
- Supplied withBoiler with water level indicator and pressure gauge, moisture trap on chamber discharge, separate steam release valves for chamber and jacket, side drain valve, high-grade stainless steel tray
- Power220/230 V or 440 V, 3 phase, 50 Hz AC
- Outer bodyClassic: SS 202 on black powder coated MS stand. Prime: all SS 304, stand included
- ConstructionInner and outer chamber, steam jacket and boiler in heavy gauge stainless steel, argon arc welded

