Wall-mounted online monitor that measures pH, temperature, dissolved oxygen and conductivity continuously and in real time, each on its own bright red LED display. Automatic temperature compensation for every parameter, and selectable units: °C or °F, ppm or % for DO, and conductivity, TDS or salinity. High and low set points for each parameter, with a built-in audible alarm and acknowledgement. Software calibration from the front keys; 3-point pH calibration with automatic recognition of pH 4.00, 7.00 and 9.20 buffers. Supplied with a pH electrode, temperature probe, DO probe, conductivity cell and buffer capsules. 12 V DC adaptor; LAN port optional.
Downloads
The Hintron HTN-MU003 watches four water-quality parameters at once and shows each one live, so a tank, line or effluent point can be monitored without taking samples to the lab. Each parameter has its own high and low set points; the built-in buzzer sounds when a reading leaves its band and stays on until it is acknowledged.
Measurement
- pH: 0 to 14.00 pH, resolution 0.01 pH, accuracy ±0.02 pH
- Temperature: 0.0 to 100.0 °C (32 to 212 °F), resolution 0.1 °C or °F, accuracy ±0.5 °C
- Dissolved oxygen: 0 to 20.0 ppm or 0 to 99.9 % saturation, resolution 0.1 ppm or 0.1 %, accuracy ±0.2 ppm or ±0.2 %
- Conductivity: 0 to 2000 µS/cm, resolution 1 µS/cm, accuracy ±2 µS/cm; TDS 0.5 % of EC and salinity 0.7 % of EC
- Automatic temperature compensation for every parameter
Calibration
- pH: 3-point, automatic or manual, with automatic recognition of pH 4.00, 7.00 and 9.20 buffers
- Temperature: 2-point (zero and span)
- Dissolved oxygen: 2-point (zero and span)
- Every parameter is calibrated in software from the front keys
Alarms and controls
- High and low set points for each parameter
- Built-in audible buzzer, with acknowledgement
- 4 soft-touch keys: PRG, INC, DEC and ENT
- Selectable units: °C or °F; ppm or % saturation for DO; conductivity, TDS or salinity
General
- Four 4-digit red 7-segment LED displays, one per parameter
- Robust ABS cabinet for wall mounting
- Power: 12 V DC, 2 A power adaptor
- Data output: LAN port (optional, at extra cost)
- Warranty: 12 months
Supplied with
- pH electrode
- Conductivity / TDS / salinity cell
- Dissolved oxygen probe
- Temperature probe
- pH buffer capsules
- Operation manual
For water-quality and process monitoring. Not for clinical use.
Measurement
- CalibrationpH 3-point, automatic or manual · Temperature 2-point (zero and span) · DO 2-point (zero and span) · Conductivity in software
- Range0 to 14.00 pH · 0.0 to 100.0 °C (32 to 212 °F) · 0 to 20.0 ppm or 0 to 99.9 % · 0 to 2000 µS/cm
- Resolution0.01 pH · 0.1 °C or °F · 0.1 ppm or 0.1 % · 1 µS/cm
- Accuracy±0.02 pH · ±0.5 °C · ±0.2 ppm or ±0.2 % · ±2 µS/cm
- Temperature compensationAutomatic · — · Automatic · Automatic
- Buffer recognitionAutomatic for pH 4.00, 7.00 and 9.20 · — · — · —
- InputCombination pH electrode (supplied) · Temperature probe (supplied) · Dissolved oxygen probe (supplied) · Cell for conductivity, TDS and salinity (supplied)
- Selectable unitspH · °C or °F · ppm or % saturation · µS/cm, TDS or salinity
- TDS and salinityDerived from conductivity: TDS 0.5 % of EC, salinity 0.7 % of EC
General
- Form factorWall-mounted
- DisplayFour 4-digit red 7-segment LED displays, one per parameter
- Power12 V DC, 2 A power adaptor
- HousingRobust ABS, wall-mounted cabinet
- Controls4 soft-touch keys (PRG, INC, DEC, ENT)
Data and connectivity
- InterfaceLAN port (optional, at extra cost)
- AlarmsHigh and low set points for each parameter; built-in buzzer with acknowledgement