Guide · Moisture and water content

Moisture analyzer: infrared or halogen balance, or Karl Fischer?

There are two ways to find out how much water a sample holds. A moisture balance heats the sample and weighs what it loses; a Karl Fischer titrator reacts the water with a reagent and counts it. This guide shows which one fits your samples, your limits and your budget.

Moisture balance
Loss on drying, 0–100 %
Karl Fischer
Water only, % to ppm
Hintron models
IR balance; 2 KF titrators
Quotation
With GST and delivery terms
Choose

Quick chooser

If you need…Choose
A fast routine moisture % for grains, powders, food or chemicalsMoisture balance (infrared or halogen)
Water only, not solvents or other volatilesKarl Fischer
Low water content: below about 0.1 %, or ppm in oils, solvents or APIsKarl Fischer
A pharmacopoeia “water content” testKarl Fischer
A pharmacopoeia “loss on drying” test, or a quick in-process checkMoisture balance
A simple method with no reagentsMoisture balance
Loss on drying

How a moisture balance works

01

Spread the sample

The sample is spread thinly on the pan of a built-in balance.

02

Heat it

A lamp above the pan heats it, so the water and other volatiles evaporate.

03

Read the loss

The balance follows the weight as it falls. When it stops changing, the loss is shown as a % of the starting weight.

Infrared balanceHalogen balance
HeaterInfrared lamp or elementHalogen lamp (a tungsten filament in a halogen-filled glass tube)
Warm-upSlowerFaster: it reaches the drying temperature quickly
Temperature controlSimplerFiner, so repeat results are closer
Typical resultsGood for routine % moistureBetter for short drying times and tight limits
CostLowerHigher
Good forGrains, cereals, powders, chemicals, food: routine QCHigh-throughput QC, heat-sensitive samples on gentle programs

Hintron HTN-MB001 is an infrared balance. It weighs and dries the sample at the same time, reads 0–100 % moisture on a net-weight basis to about ±0.5 %, and has a heat/light controller, a built-in thermometer and an analogue dial compensated 0–100 %. It runs on 220 V AC, 50 Hz.

Karl Fischer

How Karl Fischer titration works

KF reagent contains iodine, sulphur dioxide, a base and an alcohol. Iodine reacts with water in a fixed ratio, so each drop of reagent used stands for a known amount of water.

01

Dissolve

The sample is dissolved in methanol in the titration vessel.

02

Titrate

The reagent is added while the vessel is stirred.

03

Find the end point

A dual platinum electrode senses the end point, when free iodine first stays in the solution.

04

Calculate

The volume of reagent used gives the water content.

Volumetric or coulometric? Hintron’s titrators are volumetric: they dispense the reagent from a burette or a pump. Coulometric KF makes the iodine electrically and is used for very low water (ppm) in liquids and gases.

HTN-KF001 Auto KF: auto-zero glass burette (10 ml) with a 500 ml reservoir, a 10-LED array showing the titration level, a built-in stirrer and a dual platinum electrode.

HTN-KF002 Microprocessor KF: motorised reagent dispensing with no glass burette, read to 0.01 ml on a backlit 16 × 2 LCD; pump calibration from the front panel, memory of reagent used, variable-speed stirrer, programmable delay with buzzer and motorised washing of the tubing. A printer is available at extra cost.

Compare

Moisture balance vs Karl Fischer

Moisture balance (IR or halogen)Karl Fischer titrator
What it measuresEverything that evaporates: water plus solvents, alcohol and other volatilesWater only, including bound water and water of crystallisation
PrinciplePhysical: weight loss on heatingChemical: reaction of water with iodine
RangeAbout 0.1 % to 100 %Volumetric: about 0.01 % to 100 %; coulometric: down to ppm
Best accuracy atMedium and high moistureLow moisture and trace water
Time per testMinutes; depends on the sample and drying temperatureA few minutes once the reagent is standardised
Sample sizeA few grams, spread thinUsually well under a gram
Sample typesSolids, powders, grains, pastes, slurriesLiquids, soluble solids, oils, solvents; insoluble solids need a KF oven or extraction
Heat-sensitive samplesCan decompose or scorch, which reads highNo heating, so no decomposition
ConsumablesNone: just powerKF reagent, dry methanol and a desiccant
Operator skillBasicTrained: reagent standardisation and handling
Running costVery lowReagent cost for every test
SafetyA hot lamp and panReagent and methanol are toxic: use a fume hood and handle with care
Pharmacopoeia testsLoss on dryingWater determination
Typical usersGrain and flour mills, food, feed, plastics, chemicals, textiles, in-process QCPharma (APIs, tablets), oils and transformer oil, solvents, petrochemicals, fertilisers

Why the results differ: a moisture balance counts every volatile as moisture, so on samples with solvents or flavours it reads higher than KF. If your specification says “water content”, use KF; if it says “moisture” or “loss on drying”, use a balance.

Using both: many labs check a balance method against KF once, then use the faster balance for routine work.

Price

What decides the price

Moisture balance: the heater (IR or halogen), readability and capacity, the display (analogue dial or digital), the temperature range, and drying programs and data output.

KF titrator: the dispensing (manual or auto burette, or a motorised pump), the display and memory, a printer, and volumetric or coulometric.

Running cost: a balance has almost none; KF needs reagent for every test, so count the reagent at your sample load as well as the instrument.

Questions

Frequently asked

What is the difference between a moisture analyzer and Karl Fischer titration?

A moisture analyzer heats the sample and measures the weight lost, which includes all volatiles. Karl Fischer measures water only, by a chemical reaction, and works down to ppm levels.

Is a halogen moisture analyzer better than an infrared one?

A halogen balance heats up faster and controls temperature more finely. An infrared balance costs less and suits routine moisture % checks.

What is the principle of Karl Fischer titration?

Iodine in the KF reagent reacts with water in a fixed ratio. The reagent used up to the end point gives the water content.

Why does my moisture balance read higher than Karl Fischer?

The balance also counts solvents and other volatiles that evaporate, and some samples decompose on heating. KF counts water only.

Which method does pharma use?

Both. Pharmacopoeia “loss on drying” tests use drying, and “water determination” tests use Karl Fischer.

Request a quotation

Not sure which method fits?

Tell us your sample, your moisture or water limit and your tests per day. We will recommend the right method and send a detailed quotation.

+91 74049 62433
Opens WhatsApp with these details filled in.
Newsletter

Product updates from Hintron

New instruments, application notes and calibration tips for laboratories, by e-mail. Unsubscribe anytime.

We send a confirmation e-mail first. See Privacy policy.

Items have been added to cart.
One or more items could not be added to cart due to certain restrictions.
Added to cart
Quantity updated
- An error occurred. Please try again later.
Deleted from cart
- Can't delete this product from the cart at the moment. Please try again later.