Moisture analyzer: infrared or halogen balance, or Karl Fischer?
There are two ways to find out how much water a sample holds. A moisture balance heats the sample and weighs what it loses; a Karl Fischer titrator reacts the water with a reagent and counts it. This guide shows which one fits your samples, your limits and your budget.
- Moisture balance
- Loss on drying, 0–100 %
- Karl Fischer
- Water only, % to ppm
- Hintron models
- IR balance; 2 KF titrators
- Quotation
- With GST and delivery terms
Quick chooser
| If you need… | Choose |
|---|---|
| A fast routine moisture % for grains, powders, food or chemicals | Moisture balance (infrared or halogen) |
| Water only, not solvents or other volatiles | Karl Fischer |
| Low water content: below about 0.1 %, or ppm in oils, solvents or APIs | Karl Fischer |
| A pharmacopoeia “water content” test | Karl Fischer |
| A pharmacopoeia “loss on drying” test, or a quick in-process check | Moisture balance |
| A simple method with no reagents | Moisture balance |
How a moisture balance works
Spread the sample
The sample is spread thinly on the pan of a built-in balance.
Heat it
A lamp above the pan heats it, so the water and other volatiles evaporate.
Read the loss
The balance follows the weight as it falls. When it stops changing, the loss is shown as a % of the starting weight.
| Infrared balance | Halogen balance | |
|---|---|---|
| Heater | Infrared lamp or element | Halogen lamp (a tungsten filament in a halogen-filled glass tube) |
| Warm-up | Slower | Faster: it reaches the drying temperature quickly |
| Temperature control | Simpler | Finer, so repeat results are closer |
| Typical results | Good for routine % moisture | Better for short drying times and tight limits |
| Cost | Lower | Higher |
| Good for | Grains, cereals, powders, chemicals, food: routine QC | High-throughput QC, heat-sensitive samples on gentle programs |
Hintron HTN-MB001 is an infrared balance. It weighs and dries the sample at the same time, reads 0–100 % moisture on a net-weight basis to about ±0.5 %, and has a heat/light controller, a built-in thermometer and an analogue dial compensated 0–100 %. It runs on 220 V AC, 50 Hz.
How Karl Fischer titration works
KF reagent contains iodine, sulphur dioxide, a base and an alcohol. Iodine reacts with water in a fixed ratio, so each drop of reagent used stands for a known amount of water.
Dissolve
The sample is dissolved in methanol in the titration vessel.
Titrate
The reagent is added while the vessel is stirred.
Find the end point
A dual platinum electrode senses the end point, when free iodine first stays in the solution.
Calculate
The volume of reagent used gives the water content.
Volumetric or coulometric? Hintron’s titrators are volumetric: they dispense the reagent from a burette or a pump. Coulometric KF makes the iodine electrically and is used for very low water (ppm) in liquids and gases.
HTN-KF001 Auto KF: auto-zero glass burette (10 ml) with a 500 ml reservoir, a 10-LED array showing the titration level, a built-in stirrer and a dual platinum electrode.
HTN-KF002 Microprocessor KF: motorised reagent dispensing with no glass burette, read to 0.01 ml on a backlit 16 × 2 LCD; pump calibration from the front panel, memory of reagent used, variable-speed stirrer, programmable delay with buzzer and motorised washing of the tubing. A printer is available at extra cost.
Moisture balance vs Karl Fischer
| Moisture balance (IR or halogen) | Karl Fischer titrator | |
|---|---|---|
| What it measures | Everything that evaporates: water plus solvents, alcohol and other volatiles | Water only, including bound water and water of crystallisation |
| Principle | Physical: weight loss on heating | Chemical: reaction of water with iodine |
| Range | About 0.1 % to 100 % | Volumetric: about 0.01 % to 100 %; coulometric: down to ppm |
| Best accuracy at | Medium and high moisture | Low moisture and trace water |
| Time per test | Minutes; depends on the sample and drying temperature | A few minutes once the reagent is standardised |
| Sample size | A few grams, spread thin | Usually well under a gram |
| Sample types | Solids, powders, grains, pastes, slurries | Liquids, soluble solids, oils, solvents; insoluble solids need a KF oven or extraction |
| Heat-sensitive samples | Can decompose or scorch, which reads high | No heating, so no decomposition |
| Consumables | None: just power | KF reagent, dry methanol and a desiccant |
| Operator skill | Basic | Trained: reagent standardisation and handling |
| Running cost | Very low | Reagent cost for every test |
| Safety | A hot lamp and pan | Reagent and methanol are toxic: use a fume hood and handle with care |
| Pharmacopoeia tests | Loss on drying | Water determination |
| Typical users | Grain and flour mills, food, feed, plastics, chemicals, textiles, in-process QC | Pharma (APIs, tablets), oils and transformer oil, solvents, petrochemicals, fertilisers |
Why the results differ: a moisture balance counts every volatile as moisture, so on samples with solvents or flavours it reads higher than KF. If your specification says “water content”, use KF; if it says “moisture” or “loss on drying”, use a balance.
Using both: many labs check a balance method against KF once, then use the faster balance for routine work.
What decides the price
Moisture balance: the heater (IR or halogen), readability and capacity, the display (analogue dial or digital), the temperature range, and drying programs and data output.
KF titrator: the dispensing (manual or auto burette, or a motorised pump), the display and memory, a printer, and volumetric or coulometric.
Running cost: a balance has almost none; KF needs reagent for every test, so count the reagent at your sample load as well as the instrument.
Frequently asked
What is the difference between a moisture analyzer and Karl Fischer titration?
A moisture analyzer heats the sample and measures the weight lost, which includes all volatiles. Karl Fischer measures water only, by a chemical reaction, and works down to ppm levels.
Is a halogen moisture analyzer better than an infrared one?
A halogen balance heats up faster and controls temperature more finely. An infrared balance costs less and suits routine moisture % checks.
What is the principle of Karl Fischer titration?
Iodine in the KF reagent reacts with water in a fixed ratio. The reagent used up to the end point gives the water content.
Why does my moisture balance read higher than Karl Fischer?
The balance also counts solvents and other volatiles that evaporate, and some samples decompose on heating. KF counts water only.
Which method does pharma use?
Both. Pharmacopoeia “loss on drying” tests use drying, and “water determination” tests use Karl Fischer.
Not sure which method fits?
Tell us your sample, your moisture or water limit and your tests per day. We will recommend the right method and send a detailed quotation.